,,,, "Test Results for the Thermal Sensors (To, T1, T2, T3, T4)",,,, "Date: 15 Jun 2005",,,, ,,,, "Tested with applied voltage of 12.09 VDC and measured the current.",,,, "These tests were performed twice, and the results stayed the same.",,,, "Sensors give 1 microamp / K, so we should see about 300 microamps for a good sensor.",,,, ,,,, "Thermal Sensor","Location","Location Description","Inside Crystal (Female pins*)","Outside Crystal (Male pins*)" "T0","B83","Bottom","0 A","0 A" "T1","B87","Beam Left","0 A","0 A" "T2","B88","Top","298.7 microAmps","0 A" "T3","B82","Beam Right","0 A","0 A" "T4","C118","Top / Upstream","298.7 microAmps","297.5 microAmps" ,,,, ,,,, ,,,, "* These are the pins on the connectors on the sphere, not on the test connector.",,,, ,,,, "Note: The LEMO connectors come in sizes ""Grosse 0, 1, and 2.",,,, "These use Grosse 0, Isolator mit 4 kontakte ",,,, "In the magazin at PSI they are listed in book 44.0021",,,,