Test of Cosmic Veto House Scintillator Material

Introduction

        Several pieces of the scintillator material to be used in the side walls of the cosmic veto house were tested to determine their response to cosmic signals.  The results are quoted as the number of photoelectrons produced in a Burle Industries S83062E 1 inch photomultiplier tube per MeV of energy deposited in the scintillator.  Data and results can be found at the end of this page in Table 1.

Setup

       The setup was simple.  All data was taken with the phototube coupled to the end of the scintillator plank via airgap.  The phototube anode was maintained at a constant -1280 V.  To collect cosmic data, a small tagging detector was placed on either side of the plank about 1/3 of the total length away from the phototube.  This allowed us to collect data from cosmics passing only through a small area of the scintillator (~100 cm square).  A simultaneos signal in the tagging detectors triggered a 200 ns gate during which the signal from the phototube was integrated.  Event rates were roughly 0.75 Hz.  The integrated signals of several thousand events were binned into a histogram via ADC and the inhouse QVT software.  Data was also compiled for signals from a pulsed LED.  The 1 kHz driving pulse was split with one output triggering a 200 ns gate and the other driving the LED (after proper attenuation to adjust the intensity of the light).  The LED was placed directly on the scintillator and oriented perpendicular to the surface at the location of the tagging detectors.  LED Data was recorded exactly as the cosmic data; the signal from the phototube being integrated while the gate was "on" and results recorded in a histogram.  Spectra for five different LED intensities were recorded.  Figure 1  shows a typical pedestal corrected cosmic ray spectrum.   For comparison figure 2 shows cosmic data taken with a wavelength shifter coupled to the scintillator via airgap and the phototube coupled to the end of the wavelength shifter.  Note that the distribution is shifted towards zero but that the pedestal is still clearly distinct from the signal.  A typical LED spectrum is depicted in figure 3.  The measurements described above were repeated for each of nine different scintillator pieces.

 

Fig.1  A typical cosmic event spectrum

 
 
 

Fig. 2  Cosmic spectrum viewed through a wavelength shifter.

 
 
 

Fig. 3  A typical LED spectrum

Analysis

      Data was analyzed via two different methods yielding two significantly different results (see Table 1).  Neither method is entirely satisfactory and perhaps the results should be considered as upper and lower bounds to the correct values.  Evidently, the data contains noise from an unknown source.

Gaussian Analysis

      This data is typically analyzed by fitting the LED data with gaussian functions (see figure 4) and plotting the squared ratio of the mean position to sigma as a function of the mean position.  This data should form a line passing through the origin whose slope is the number of photoelectrons per ADC channel.  It can then be determined how many photoelectrons correspond to the position of the mean of the cosmic spectrum.  Since the mean energy deposited by cosmic signals is known, this in principle determines the number of photoelectrons per unit of energy deposited in the scintillator.  The width (sigma) of the distribution has contributions from two sources.  The first is the actual statistical variation in the number of photoelectrons produced.  It is this variation which we wish to quantify.  The second contribution comes from random noise.  This noise must be accounted for in the analysis.  In principle the random noise should remain constant throughout the experiment such that the measured sigmas are all a constant amount greater than the true statistical variation.   Figure 5 shows a plot of sigma squared versus the mean of the distribution for a typical piece of scintillator.  This data should form a line through the origin.  The presence of random noise of course shifts the y-intercept above the origin.  There is obviously an additional source of noise which is not constant.  Figure 6 shows the resulting plot of the squared ratio of the mean to the corrected sigma.  Note that the error bars on the data points (table 1) are again extremely small resulting in a very poor fit.  Furthermore the contribution of noise to the raw sigma used to compute these data points is itself based on a poor fit (figure 5).

Fig. 4: data with gaussian fits for one scintillator



Fig. 5 (raw sigma)^2 versus mean for a single scintillator

 
 
 

Fig.6  (mean/corrected sigma)^2 versus mean for a single scintillator.

     A quick glance at Table 1 (and figure 4) shows that the gaussian fits are quite good (chi^2/DOF ~ 1) and the statistical errors on the fit parameters are very small.  However, the linear fits discussed above are very poor.  As a solution to this problem we keep only the three data points with the lowest means.  The means of the cosmic spectra are well contained within this range of means for each scintillator and the three remaining points form a more reasonable line (see Figure 7).   The plots discussed so far display data for a single scintillator plank (#322).  Table 1 shows the complete data and results from all tested planks.  The final result for the number of photoelectrons per MeV of energy is the mean of the results for each individual piece and the error quoted is the statistical error in the mean.

 Fig. 7 (mean/corrected sigma)^2 versus mean for only the 3 points with the lowest means.

Poisson Analysis

      As an alternative method of analysis we find the best fitting Poisson distribution for each led spectrum (see figure 8).  In this approach the fit parameters are the number of ADC channels per photoelectron and the mean of the distribution.  This method clearly provides a more direct path to the desired result as one of the fit parameters is the gain of the photomultiplier/ADC combination and a close inspection of figure 9 reveals that Gaussian statistics may be inappropriate anyway (note a slight asymmetry between the left and right tails).  The drawback to this method is that there is no way to account for noise as with the Gaussian analysis.

Fig 8  Led data and corresponding poisson fits (data in solid lines, fits in dashes) for a single piece of scintillator.  Note that the scales are different since data must be rebinned to match the number of ADC channels per photoelectron in the trial Poisson distribution.

Fig. 9  As the intensity of LED pulses is decreased the spectrum begins to deviate from a Gaussian

     We obtained the number of ADC channels per photoelectron from the fit of each LED spectrum.  This number is the fit parameter yielding the best value of chi^2.  Comparison with the corresponding cosmic distribution yields the number of photoelectrons corresponding to the mean of the cosmic distribution (measured in ADC channels).  The dimensions of the scintillator and dE/dx for cosmic muons in the scintillator material is known so that we know the mean energy deposited in the material (~ 4.5 MeV).  This information allows one ultimately to compute the mean number of photoelectrons produced in the phototube per MeV of energy deposited in the scintillator.  A value was computed for 5 data points in each of 9 planks and the results averaged to give the final result (see table 1).

Scintillator Efficiency

       The efficiency of each scintillator piece was also computed.  The efficiency is defined as (signal events - pedestal events)/(total events) for cosmic data.  All pieces are more than 98% efficient.

 

 
Gaussian Fits Poisson Fits
Scintillator ID Cosmic mean mean sigma^2 chi^2 (mean/corrected sigma)^2 electrons/MeV ADC ch/e- chi^2 electrons/MeV
000* 416.17 158.62(0.18) 1407.00(9.75) 1.49 47.55(0.03) 27.51 9 2.39 10.27
226.61(0.18) 1643.49(10.50) 1.11 64.72(0.04) 8 2.02 11.56
258.04(0.19) 1733.89(11.66) 1.02 77.79(0.04) 7 1.07 13.21
335.74(0.20) 1904.45(12.22) 2.20 109.81(0.05) 6 1.10 15.41
516.10(0.20) 2037.64(12.64) 1.41 264.63(0.07) 6 41.64 15.41
227 247.44 126.93(0.16) 1327.14(8.01) 4.09 35.12(0.02) 14.77 11 4.12 5.00
171.15(0.18) 1526.46(10.16) 0.93 44.52(0.03) 9 2.78 6.10
219.77(0.17) 1663.82(10.60) 1.42 60.73(0.04) 8 1.44 6.87
281.35(0.19) 1884.43(11.29) 0.93 77.92(0.04) 7 1.22 7.85
346.77(0.20) 1998.98(12.52) 0.99 106.37(0.05) 6 1.09 9.16
228 239.38 147.01(0.16) 1431.87(8.32) 1.51 42.95(0.02) 15.45 10 3.61 5.32
196.01(0.17) 1612.02(10.44) 0.94 56.23(0.03) 8 2.53 6.65
252.93(0.18) 1797.76(11.02) 1.09 73.61(0.04) 7 1.31 7.60
317.91(0.18) 1940.40(11.45) 1.08 99.90(0.04) 6 1.62 8.87
390.54(0.16) 2030.40(10.81) 1.45 138.44(0.05) 5 2.36 5.32
3110 241.97 136.32(0.17) 1480.71(9.24) 1.46 35.64(0.02) 13.78 11 4.50 4.89
183.30(0.15) 1694.97(9.06) 1.30 45.67(0.02) 10 4.45 5.38
237.16(0.19) 1875.76(11.26) 1.35 61.37(0.03) 8 1.74 6.72
317.91(0.18) 1940.40(11.45) 1.08 103.01(0.04) 7 1.53 7.68
370.44(0.19) 2190.24(13.10) 0.90 111.48(0.05) 6 1.28 8.96
317 293.56 111.68(0.17) 1329.33(8.75) 2.54 21.89(0.02) 12.37 13 4.07 5.02
152.04(0.17) 1597.00(9.59) 1.20 27.60(0.02) 11 3.65 5.93
200.82(0.18) 1802.85(11.04) 2.42 38.66(0.03) 9 2.98 7.25
317.91(0.18) 1940.40(11.45) 1.08 85.59(0.04) 8 3.11 8.15
320.55(0.18) 2202.43(12.20) 1.25 71.22(0.04) 7 2.03 9.32
321 252.14 142.99(0.16) 1407.00(8.25) 1.85 43.68(0.02) 16.57 10 3.08 5.60
190.79(0.18) 1615.24(10.45) 1.13 53.82(0.03) 9 2.00 6.23
246.25(0.18) 1758.96(10.90) 1.23 73.94(0.04) 7 1.04 8.00
310.88(0.18) 1936.00(11.44) 1.09 96.93(0.04) 6 1.50 9.34
381.58(0.19) 2033.11(12.63) 1.13 133.07(0.05) 5 2.01 11.21
322 228.83 141.49(0.17) 1393.53(8.96) 1.73 35.33(0.03) 12.62 10 3.63 5.09
189.58(0.17) 1596.00(9.59) 2.26 46.73(0.03) 9 3.23 5.65
243.03(0.19) 1801.15(11.03) 0.99 60.62(0.04) 8 2.06 6.36
307.17(0.19) 1920.19(11.39) 1.13 86.30(0.04) 6 1.69 8.48
377.02(0.19) 2040.33(11.74) 1.05 117.14(0.05) 5 2.43 10.17
7211 279.45 128.62(0.17) 1445.66(9.15) 1.58 35.63(0.02) 16.33 12 4.11 7.76
177.42(0.17) 1702.39(9.90) 1.08 43.65(0.05) 10 4.05 10.35
234.99(0.18) 1854.16(10.33) 1.32 63.26(0.04) 8 2.97 8.87
301.70(0.18) 2025.90(11.70) 0.96 87.14(0.03) 7 1.58 8.48
374.06(0.18) 2092.15(11.89) 1.53 125.96(0.03) 6 2.09 5.18
729 334.85 114.04(0.17) 1320.60(8.90) 2.33 24.56(0.02) 15.49 13 5.46 5.72
155.08(0.17) 1569.74(9.48) 1.18 30.89(0.03) 10 3.48 7.44
203.67(0.19) 1753.93(10.88) 1.33 43.08(0.04) 9 2.21 8.27
259.88(0.18) 1939.52(11.47) 1.15 58.81(0.04) 8 2.33 9.30
323.85(0.18) 2104.06(1.85) 1.45 79.88(0.05) 7 2.23 10.63
average 14.67(1.60) 7.44(1.83)
 
* Scintillator 000 was not included in the averages as it was only half the width of the other board.