next up previous contents
Next: Outlook Up: Conclusion Previous: Experimental Method and

Detector Testing and Calibration

After delivery, the CsI crystals must pass three tests to be selected for use in the calorimeter: the ratio of the fast component to total scintillation light output must be greater than seventy percent; the linear and angular dimensions of a given crystal must be within predefined tolerances; the light collection non-uniformity must be better than three percent. Various surface treatment methods have been investigated for maximal light output and maximum non-uniformity and a cosmic ray tomography apparatus has been set up to measure the light non-uniformity throughout the volume of each module. A particular surface treatment consisting of three layers of teflon plus one layer of aluminized mylar was arrived at by the time of the development run of 1994; as a result, the modules used during that run were wrapped as described above. The results of the development runs aided in the optimization of the design of the apparatus, revealed the performance of the main detector systems and validated the experimental technique. During the test of 1992, the effort was concentrated exclusively on the beam and its parameters such as the spot size, the horizontal and the vertical profiles, the contaminations and the beam particle rates. The results of this test affected the design of the active target and its segmentation. The target was fabricated at the University of Virginia and inserted in the beam during the test run of 1993. By that time, some CsI modules were delivered by Bicron Corporation and it was possible to build up a section of the calorimeter and to study the responses of the array to monoenergetic beams of positrons at various momenta. The line shapes of the responses of the array were compared to simulations. Excellent agreement was observed after the effects of photoelectron statistics and light collection non-uniformity were accounted for. At 70 MeV/c (this is the energy of interest since the 's from have ) the measured energy resolution of the calorimeter was 5.6 MeV FWHM. The study of the stopping rate in the active target revealed that the number of stopped pions scaled with the area of twenty-five, nine and four fibers. This allowed the redesign of the active target with fibers instead of . The redesigned active target was also built in Virginia and tested in beam during the test run of 1994: the beam distributions measured with the active target suggest a beam spot of FWHM. Another study conducted during the 1993 run was a test of the wrapping method. Of the various surface treatments considered, black paper at the front faces of the crystals with the lateral surfaces wrapped in teflon gave the best light output and uniformity. Further studies after the run led to the wrapping system used in the tomography apparatus and during the test run of 1994. In the tomography system, the directions of the cosmic muons are defined by three drift chambers. The path-lengths of the cosmic rays and their energy depositions in the CsI modules are used in a reconstruction algorithm which finds the light per unit path-length in the volumes of the modules --- the crystals are subdivided into cubic cells and the reconstructed light yields per path-lengths in the cells constitute a measure of the optical non-uniformity in the crystals. The CsI modules used during the test of 1994 were all examined in the tomography system. More crystals were delivered by that time and it was possible to set up a much bigger portion of the calorimeter than the one used in the 1993 experiment. The line shapes of the responses of the calorimeter were remeasured with not only monoenergetic 's but monoenergetic 's as well. With a good control of the temperature and the humidity, a better wrapping system, and a bigger size of the array, a better energy resolution was achieved, i.e., 4.2 MeV FWHM. In addition, the response of the array to decays confirmed the need for MWPC's with good double track resolution for additional suppression factors of double Michel events. The decays were well separated from the Michel background and the number of events collected was consistent with the correct branching ratio.


next up previous contents
Next: Outlook Up: Conclusion Previous: Experimental Method and



Bernward Krause
Mon Jan 15 14:57:06 MET 1996