S121-1a TOMOGRAPHY PAGE

S121-1a Surface treatment 1a: Runs 620-662

Axial Light Output Uniformity

Transverse Light Output Uniformity

2D Light Output Uniformity

Coordinate offset x_off: 45.90 cm, y_off: 29.80 cm
Fast:Total= 0.753
Photoelectrons/MeV= 56
Temperature Coefficients (f,t,d,f:d,f:t): -0.52,0.27,3.40,-4.04,-0.88 %/deg C
Axial nonuniformity slopes (A1,A2,A3)= -0.89,-2.50,0.50 %/cm
Transverse nonuniformity slopes (T3L,T3R,T18L,T18R)= %/cm
Timing Resolution= ns

The axial nonuniformity is calculated as follows:

Light output per unit pathlength in the crystal (s/d) is plotted as a function of z-position in the crystal. The front face of the crystal corresponds to z=0 cm, while the back of the crystal is z=22 cm. The nonuniformity constant is equal to the slope of this scatter plot, divided by the average value of s/d in the region of interest. Using this definition, a negative axial nonuniformity corresponds to a decreasing value of s/d from z=0 to s=22 cm.

The transverse nonuniformity is calulated as follows:

"s/d" is plotted as a function of x-position in the crystal. Looking at the back face of the crystal, the positive x-axis points to the right. The nonuniformity constant is equal to the slope of the scatter plot, divided by the average value of s/d in the region of interest. A negative transverse nonuniformity on the right side of the crystal corresponds to a decreasing value of s/d from the center to the edge of the crystal, while a positive transverse nonuniformity on the left side of the crystal means that s/d increases toward the center.

E. Frlez, 11 Mar 1996