By assuming that the target is a point source, the angle of incidence of the particle can be calculated for each event. This approximation holds as long as the actual target size is small compared to the distance between the target and detector arrays.
For the NaI detector, the angle of incidence correction is calculated along the two directions in the plane of the front face of the array, and . The technique used is similar to the one suggested by A. Bay et. al. [1]. The shift (or ) is applied according to Fig. 5.2, where is related to the electromagnetic shower penetration depth and is a function of energy. From Fig. 5.2 one can see that .
Figure 5.2: Geometry used to calculate angle of incidence corrections
to shower reconstruction coordinates in NaI array.
For the CsI detector, which has a spherical geometry, the angular
correction is applied to produce a shift in the and
directions, which are also along the front face of the array.
First, the angle of incidence with the normal to the detector surface
is calculated in each direction and , as shown in
Fig. 5.3.
Next, because the linear correction x is small compared to the distance
to the target r, one
can assume that . Assuming that the linear
correction x is much less than the sphere radius R, one can see from
Fig. 5.3 that
.
Finally, the angular correction
in both the and directions takes the form
Figure 5.3: Geometry used to calculate the angle of incidence correction
to the reconstructed shower coordinates in the CsI array.
The values of the shower depth parameters and have been chosen for optimal average angular resolution. Figure 5.4 shows the plots of average angular deviations plotted against , for 55-83 MeV photons from the reaction in the GEANT simulation. One can see that the optimal values are cm and cm. These penetration depths reflect the fact that the NaI crystals have a lower atomic number than the CsI crystals, shown in the larger value for .
Figure 5.4: Average angular deviation plotted against the shower
penetration depth parameter , for the CsI array (top panel)
and NaI array (bottom panel).
Finally, the event-by-event angular deviations are plotted in Fig. 5.5, with and without the angle of incidence corrections. It is clear that there is a significant improvement with the inclusion of the correction. The CsI decreases from 2.3 to 1.8, and the NaI from 1.3 to 1.2.
Figure 5.5: Angular deviations , with and without the angle of
incidence corrections, for the CsI array (top panel) and NaI array
(bottom panel).